Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.
Mirion Technologies (MGPI) SA
Route d’Eyguières
FR-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna, GA 30082
USA
T +1 770 432 2744
F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a
DE-80687 Munich
Germany
T +49 (0) 89515 13-0
F +49 (0) 89515 13 169
Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu
Zhabei District, Shanghai 200436
PR of China
T +86 21 6180 6920
F +86 21 6180 6924
NGM 216S
Particulate, Gas & Iodine Monitors
PHYSICAL CHARACTERISTICS
● Radiation detected: beta
● Detector: 2’’ thin plastic scintillator + PMT +
embedded LED
● Lead shield: 4 π/7.5 cm (4 π/3 in)
● Typical energy range: > 30 keV
● Typical measurement range: 3.7 10
+3
to 3.7 10
+9
Bq/m
3
(10
-7
to 10
-1
μCi/cc)
ENVIRONMENTAL CHARACTERISTICS
● Long term temperature: +10°C to +40°C
(+50°F to +104°F)
● Periodic temperature (processing unit):
-5°C to +55°C (+23°F to +131°F)
● Periodic temperature (detector):
+0°C to +60°C (+32°F to +140°F)
● MTBF (processing unit): > 50 000 hours
● MTBF (detector): > 20 000 hours, with
preventive maintenance
● TID: 100 Gy (10
+4
rad)
● Protection index: IP65 and IK07
PNEUMATIC CHARACTERISTICS
● Standard ow rate: 28.3 l/min (1 scfm)
● Pressure drop: 50 mbar (0.73 psi)
MECHANICAL CHARACTERISTICS
● Dimensions:
○ Processing unit: 390 mm x 196 mm x 187 mm
(15.3 in x 7.7 in x 7.3 in)
○ Detector: 250 mm (10 in) x Ø 72 mm (2.8 in)
● Weight:
○ Processing unit: 4.5 kg or 7 kg (10 lb or 15.5 lb)
○ Detector: ~ 300 kg (661 lb)
● Color: gray RAL 7030 (decontaminable paint)
● Inlet tube connection: Ø 12 mm OD (1/2 in)
● Outlet tube connection: Ø 12 mm OD (1/2 in)
ELECTRICAL CHARACTERISTICS
● Power supply: 230 Vac – 50 Hz or 120 Vac – 60 Hz
● Data link outputs: 1 RS232 and 2 isolated RS485
● Alarm relays: 3 SPDT relays
● I/O: 2 isolated analog outputs (0/4-20 mA)
● Embedded LED tester
SIGNALING (Applicable to LPDU only)
● Alphanumeric display: measurement, status…
● Sound alarm: buzzer 90 dBA at 1 meter
● Visual alarm: 3 lights (red, yellow, green)
REFERENCE STANDARDS
● Nuclear: IEC60761-1 and IEC60761-3
● Environmental: IEC60780, IEEE323
● Seismic: IEC60980, IEEE344
● EMC: 2006/95/CE and 2004/108/CE, EPRI 102323,
MIL STD 461 E, IEC61000-6-2 and IEC61000-6-4
VERSIONS
● 230 Vac or 120 Vac
● LPDU or LPU
● With or without check source
● With or without heater
● PIS sampler
● Dust lter holder
ACCESSORIES
● Calibration tools
● Software
● Ethernet
● USB converters
www.mirion.com